W-2532
This is an error code that comes up periodically on the Frontier 570.
Warning W-2532
G laser (G-SHG) optimal temperature setting periods.
Press [Help] button to see error help (or manual). Set optimal temperature for G laser (G-SHG) while referring to it.
Course of Action
- In the admin-Panda application on the 570 computer, click on 4-Maintenance on the lower part of the application window.
- A window will open up. Type the password 7777 and click enter. This will let you enter the application as an administrator.
- Click on 5-Printer Adjustment/Maintenance.
- Then click on item [454] - G Laser(SHG) Optimal Temperature Setup.
- Click Enter. This will cause the system to perform the laser optimal temperature adjustment. The process will take about 5 minutes.
- Once the process it finished the Measured value should read OK.
- Click Close. This will take you back to the admin-Panda and you'll be ready to start printing.